• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXTEND; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXTEND; IMPLANTABLE LEAD Back to Search Results
Model Number 4086
Medical Device Problem Codes Failure to Capture (1081); Over-Sensing (1438); Under-Sensing (1661)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a request was made for technical services (ts) to review stored atrial tachy response (atr) episodes for this patient.Upon review, ts found evidence of undersensing, oversensing and possible loss of capture (loc) of this right atrial (ra) lead.Ts considered lead integrity issues and lead dislodgement as possible causes.According to the health care professional (hcp), a recent chest radiograph showed atrial lead in proper position.Ts recommended in person evaluation for this patient.No adverse patient effects were reported.This ra lead remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTEND
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23722506
Report Number2124215-2025-89259
Device Sequence Number18351703
Product Code DTB
UDI-Device Identifier00802526397721
UDI-Public(01)00802526397721(17)150624
Combination Product (Y/N)Y
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
P960006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/24/2015
Device Model Number4086
Device Catalogue Number4086
Device Lot Number263187
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Initial Report FDA Received Date12/04/2025
Date Device Manufactured06/24/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
-
-