| Model Number |
18870 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: 18870, 18870 7.0mm taperguard evac trachealx10, (lot# 24i0210jzx) 18875, 18875 7.5mm taperguard evac trachealx10, (lot# 25d1092jzx) 18875, 18875 7.5mm taperguard evac trachealx10, (lot# 25d1092jzx) 18875, 18875 7.5mm taperguard evac trachealx10, (lot# 25d1092jzx) 18875, 18875 7.5mm taperguard evac trachealx10, (lot# 25d1092jzx) 18875, 18875 7.5mm taperguard evac trachealx10, (lot# 25d1092jzx).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the cuffs of the seven endotracheal tubes had poor sealing prior to use.During the test, the cuff pressure was maintained at 25-30 cmh2o, but liquid leakage was still observed at the top of the cuff, indicating poor cuff seal.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that that the cuff of the endotracheal tube had poor sealing prior to use.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|