A non-healthcare professional reported that during an intraocular lens (iol) implant procedure, the haptic does not unfold, remains straight, with no patient contact.The procedure was completed on the same day without any harm to the patient.
|
|
Additional information has been provided in d.9.,h.3., h.6., and h.11 the device with the lens was returned in the carton.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was at the trailing optic edge.The lens was advanced to mid-nozzle.The trailing haptic was folded on the optic.The leading haptic was extended.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A qualified viscoelastic was used.No problems were observed with the returned device.The plunger, lens and haptic positions were acceptable.The leading haptic was straight.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the ifu (instructions for use).A straight leading haptic may occur: due to the normal folding variations as indicated in the ifu diagrams.If the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.If there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement or contribute to incorrect haptic folding.Any of the above listed causes alone, or in combination, may create the reported event.If a straight leading haptic occurs, the ifu instructs: insert the tip of the cartridge through the incision.Position the tip at the anterior capsule opening.If the leading haptic is looped or straight in front of the optic, rotate the nozzle clockwise as needed to be bevel left to facilitate placement of the leading haptic into its normal orientation in the bag.Slowly advance the lens until the optic is exiting the nozzle.Rotate the cartridge counter-clockwise towards bevel right as needed to place the lens anterior side up.The manufacturer internal reference number is: (b)(4).
|