RESPIRONICS, INC. AERIS EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VT2110X24B
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information that an aeris evo ventilator was returned to the third-party service center for completion of required maintenance.There was no harm or injury reported.Device evaluation revealed a ventilator service required error code in the download error log indicating the internal or detachable li-ion battery is not charging due to a hardware fault for greater than or equal to 1 minute.Evaluation determined the device's power supply required replacement to address this issue.Preventative maintenance was completed.The device passed final testing after repair and was determined to operate properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AERIS EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23726013
Report Number2518422-2025-112347
Device Sequence Number10642093
Product Code CBK
UDI-Device Identifier00606959062962
UDI-Public(01)00606959062962
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberVT2110X24B
Device Catalogue NumberVT2110X24B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Initial Report FDA Received Date12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-