It was reported that during an anterior cruciate ligament bump fellow bone procedure working on the femoral side, while using the milagro advance screw 9x23mm device, the tunnel and the bone block were aligned in line meaning we drilled a size 10 and inserted a size 10 bone block.Once inserting the mega wire cap was introduced and milagro screw was inserted, and crack was able.The scissor was able to remove the fragments.Another device was used to complete the procedure.The exact date of this event was unknown but was noted to have occurred in 2025.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the complaint device was discarded by the customer, therefore not returned to depuy synthes and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|