|
DEPUY MITEK LLC US MICROFIX QA+#3/0 OC V-4; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
212859 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure the microfix qa+#3/0 oc v-4 device anchor broke when the knot was tightened.Another anchor was used to complete the procedure.There was no delay in the procedure reported.The exact date of this event was unknown but was noted to have occurred in 2025.There were no adverse patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was discarded by the customer, therefore not returned to depuy synthes and unavailable for a physical evaluation.Given that no lot/serial number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot/serial number becomes available, the mre review will be performed.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|
|
|