Catalog Number 05.001.080
Medical Device Problem Code Fail-Safe Did Not Operate (4046)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
 
Event or Problem Description
It was reported that during service and evaluation, it was determined that the air pen drive (apd) device had corrosion/rusting/pitting, the device would run in the locked position, had insufficient/low power, and an air leak.It was further determined that the device failed pretest for general condition, check handpiece in ¿lock¿ mode, check general function with apd hand switch, check power with power test bench, check speed with tachometer, check for external leak tightness, and check for internal leak tightness.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR PEN DRIVE
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
elba bello
eimattstrasse 3
oberdorf 04436
SZ   04436
9083863534
MDR Report Key23727904
Report Number8030965-2025-12203
Device Sequence Number18696039
Product Code HBB
UDI-Device Identifier07611819830899
UDI-Public(01)07611819830899
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K093361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2025
Initial Report FDA Received Date12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-