| Catalog Number |
960783 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Foreign Body Reaction (1868)
|
| Date of Event |
11/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that universal fem slv dis por 31mm, sigma fem adapter neutral bolt, and an unknown knee femoral adaptor were involved in a reported event.The event concerned a tc3 depuy prosthesis where, during a post-operative period, abundant metallosis was observed in the knee without noticeable wear of the polyethylene.Upon extraction, abnormal rotational movement was detected between the femoral shield and the metaphyseal sleeve, which are typically joined by an impacted morse cone system.This movement was identified as the source of metal debris.The femoral component remained held by the keel in the femoral diaphysis, but the epiphyseal part of the implant exhibited rotational mobility relative to the femur.The neutral bolt and associated parts appeared visually sealed.The prosthesis was changed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|