• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC OSSEOTITE® TAPERED IMPLANT 4 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC OSSEOTITE® TAPERED IMPLANT 4 X 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number NT410
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
Date of Event 09/23/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implants at tooth sites #7, and #10 loss integration due to peri-implantitis.Implants were removed.Patient had pain and inflammation.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).Multiple reports are being submitted for this event.A4: patient weight unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED IMPLANT 4 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark  mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23728939
Report Number0001038806-2025-03620
Device Sequence Number21249559
Product Code DZE
UDI-Device Identifier00844868012016
UDI-Public(01)00844868012016(17)270713(10)2022070293
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K063286
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberNT410
Device Lot Number2022070293
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-