| Catalog Number |
03.404.026S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Cardiac Arrest (1762); Embolism/Embolus (4438)
|
| Date of Event |
11/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported the drain tree became fused with the tubing and burr during the reaming of the canal, and the tubing ruptured inside it.At the start of the surgery, the tubing leaked, with water escaping.The patient went into cardiac arrest, and removing the drain tree proved too complex, requiring a larger surgical approach and a femoral osteotomy.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: a2 (age), b5, h6 (health effect - clinical code) corrected: h5, h8.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received: a 43-year-old patient with a 16-year history of chronic osteomyelitis of the right femur was hospitalized on (b)(6) 2025, due to pain and a non-active fistula; at the time of the procedure, the patient was in good condition but had poor pain control, with a fistula showing reluctance, flushing, edema, and erythema, without fever or other symptoms.The surgical plan was to wash the right intramedullary canal; however, during assembly of the ria 2 system, a saline leak occurred.Milling was initiated using synream and manual cutters, followed by passage of the ria with a 15 mm milling head.During reaming, the ria hose became blocked, and the patient suffered cardiac arrest requiring cardiopulmonary resuscitation.Due to inability to remove the trapped hose, an urgent subtrochanteric osteotomy was performed.Post-procedure review revealed the shaft fused with the hose.The patient was placed on ecmo and transferred to the icu, later referred to shaio clinic in bogotá.Ecmo was discontinued after three days; brain mri revealed cerebral embolism, and the patient subsequently died at shaio clinic.
|
| |
|