Catalog Number 8065752435
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Eye Irritation (4803)
Date of Event 11/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported that beveled tip of the vitreous cutter became detached and dropped into the eye and eye discomfort during vitrectomy surgery.The surgery was completed by replacing the product on the same day.There was no patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in a.2., a.3., b.5 and h.11.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was received that the detached part was immediately removed upon detected and replaced the vitrectomy probe.
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.11.No technical inquiries were received from the customer.A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Photo attached to the file was reviewed by the manufacturing site.Photo 1 shows a probe in the eye with the tip missing and photo 2 shows a piece of foreign material circled in red on a screen.The reported issue is confirmed from the photo review.The investigation conducted a non-conformance review of the reported lot number.No deviations, were identified and all corresponding production release specifications defined in the device master record were met.The reported issue is confirmed from the photo review; however, since the sample was not returned and based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23729525
Report Number1644019-2025-05328
Device Sequence Number16105950
Product Code LRO
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752435
Device Lot Number17FLC4
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/10/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/11/2025
Supplement Date Received by Manufacturer12/11/2025
03/31/2026
09/03/2026
Initial Report FDA Received Date12/05/2025
Supplement Report FDA Received Date01/07/2026
04/20/2026
09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-