• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR; Permanent defibrillator electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR; Permanent defibrillator electrodes Back to Search Results
Model Number 7170Q/65
Medical Device Problem Codes Break (1069); High impedance (1291); Device Dislodged or Dislocated (2923); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was noted that the patient's right atrial lead exhibited noise and right ventricular (rv) lead exhibited high defibrillation impedance.During replacement procedure, it was discovered that the ra and rv leads were twisted improperly around the device and both had insulation damage.The ra and rv leads were explanted and replaced.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR
Common Device Name
Permanent defibrillator electrodes
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key23730129
Report Number2017865-2025-1006650
Device Sequence Number13197432
Product Code NVY
UDI-Device Identifier05414734503266
UDI-Public(01)05414734503266(10)S000030055(17)180228
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Model Number7170Q/65
Device Lot NumberS000030055
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2025
Initial Report FDA Received Date12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/22/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GALLANT.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
Patient Weight79 KG
-
-