BAXTER HEALTHCARE CORPORATION TRANSPORT STRETCHER; STRETCHER, WHEELED Back to Search Results
Model Number P8005G
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
During servicing by baxter, technical service identified that std stretcher frame (product code: p8005g000030, serial number: (b)(6), had brakes not holding.This occurred during baxter servicing/testing.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
 
Additional Manufacturer Narrative
The baxter technician found the casters needed to be replaced.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Apply the brake and make sure the stretcher does not move.If there is movement, look at the brake components for wear.Apply the steer, and make sure the stretcher steers correctly.Look at the steer components for wear.Put the stretcher in neutral.Make sure all four casters rotate and roll freely.Adjust or replace components if necessary.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the casters to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPORT STRETCHER
Common Device Name
STRETCHER, WHEELED
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
HILL-ROM MEXICO
ave.del telefono
no. 200 col. huinala
apodaca, nuevo leon 66640
MX   66640
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key23730385
Report Number3006697241-2025-00326
Device Sequence Number9667656
Product Code FPO
UDI-Device Identifier00887761000964
UDI-Public(01)00887761000964(11)090810
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberP8005G
Device Catalogue NumberP8005G000030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2025
Initial Report FDA Received Date12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Patient SexUnknown
-
-