• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH MATNEU SCR ø1.5 SELF-DRILL L5 TAN 4U I/C; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH MATNEU SCR ø1.5 SELF-DRILL L5 TAN 4U I/C; PLATE, BONE Back to Search Results
Catalog Number 04.503.105.04S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.
 
Event or Problem Description
It was reported two screws broke during surgery.A third screw was used to successfully complete the procedure with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9, corrected: d4 primary udi number.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 (expiry date), h4.Corrected: d4 (primary udi number), e1 (facility name), g1 (manufacturing site name).H3, h4, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that matneu scr ã¸1.5 self-drill l5 tan 4u i/c was broken and only a fragment of the screw was return.The fracture surface was thoroughly examined and no problems with the material were noted.The observed condition is consistent with a torsional failure which is likely due to the application of significant torque during the implantation process.According to the matrixneuro cranial plating system guide the following warnings are mentioned: - these devices can break during use (when subjected to excessive forces or outside the recommended surgical technique).While the surgeon has to make the final decision on removal of the broken part based on associated risks in doing so, we recommend that whenever possible and practical for the individual patient, the broken part should be removed.- be aware that implants are not as strong as native bone.Implants subjected to substantial loads may fail.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the matneu scr ã¸1.5 self-drill l5 tan 4u i/c would have contributed to the complained device issue was consistent with the reported condition.Based on the investigation findings, the potential cause is traced to the application of significant torque during implantation.There is no indication that a design or manufacturing issue has caused the reported complaint condition and it has been determined that no corrective and/or preventative action is proposed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): a manufacturing record evaluation was performed for the finished device.Product code: 04.503.105.04s-15.Lot number:65805p0.It was electronically reviewed and no nonconformances/manufacturing irregularities were identified during the manufacturing process.The product was released on: 22 may 2025.Manufacturing site: jabil bettlach.Expiry date: 01 may 2035.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATNEU SCR ø1.5 SELF-DRILL L5 TAN 4U I/C
Common Device Name
PLATE, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach CO 2544
SZ   2544
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key23731752
Report Number8030965-2025-12233
Device Sequence Number13419776
Product Code JEY
UDI-Device Identifier10887587018355
UDI-Public(01)10887587018355(17)350501(10)65295P0
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K123723
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.503.105.04S
Device Lot Number65295P0
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/18/2025
Supplement Date Received by Manufacturer12/17/2025
01/21/2026
Initial Report FDA Received Date12/05/2025
Supplement Report FDA Received Date12/19/2025
01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-