|
(b)(4) date sent: 12/5/2025 d4 batch #: y7050x.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the distal tip of the blade broken off and it was returned with the device.The remaining blade portion was scratched, with evidence of contact with metal in or out of the operative field.During functional testing on energy systems, an alert screen was displayed.A probable cause for the device to stop activating and the energy systems to display an alert screen is blade damage.The device was disassembled to verify the condition of the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage, which can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens can include "remove instrument from patient," ¿blade error detected,¿ or "relaxed pressure on blade," followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch y7050x, and no non-conformances were identified.
|