• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTSINE TECHNOLOGIES LTD BATTERY,PAD-PAK-02; AUTOMATED EXTERNAL DEFIBRILLATOR ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTSINE TECHNOLOGIES LTD BATTERY,PAD-PAK-02; AUTOMATED EXTERNAL DEFIBRILLATOR ELECTRODE Back to Search Results
Model Number PAD-PAK-02
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pad-pak was not returned to heartsine for evaluation.The customer has been instructed to dispose of the affected unit following local waste disposal regulations.The customer was sent a replacement pad-pak.A capa investigation determined multiple potential causes of the fault, both internal and external to stryker manufacturing processes, the primary root cause was determined to be pressure applied onto the locator pins during handling/packaging.
 
Event or Problem Description
In response to a product field action, the customer contacted heartsine to report the pad-pak's locator pin is bent.This issue can lead to a failure to detect when a patient is connected, or could cause the device to lose power during operation.There was no patient involvement reported with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY,PAD-PAK-02
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR ELECTRODE
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
Manufacturer (Section G)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI   BT3 9ED
Manufacturer Contact
brian blakeslee
207 airport road west
belfast BT3 9-ED
EI   BT3 9ED
4258674577
MDR Report Key23735109
Report Number3004123209-2025-00460
Device Sequence Number21473360
Product Code MKJ
UDI-Device Identifier05060167120473
UDI-Public05060167120473
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P160008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 12/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPAD-PAK-02
Device Catalogue NumberPAD-PAK-02
Device Lot NumberL0230
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2025
Initial Report FDA Received Date12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RA2025-4068245
Patient Sequence Number1
Patient SexUnknown
-
-