• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SALTER LABS N/A; NASAL CANNULA (ADULT) HIGH FLOW WITH 7' (2.1 M) SUPPLY TUBE, FOR FLOWS UP TO 15

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SALTER LABS N/A; NASAL CANNULA (ADULT) HIGH FLOW WITH 7' (2.1 M) SUPPLY TUBE, FOR FLOWS UP TO 15 Back to Search Results
Model Number 1600HF-7-25
Medical Device Problem Codes Insufficient Flow or Under Infusion (2182); Obstruction of Flow (2423)
Health Effect - Clinical Code Low Oxygen Saturation (2477)
Date of Event 10/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 04 dec 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.
 
Event or Problem Description
It was reported that o2 tubing found to have a hole in between the nasal prongs.
 
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 04 dec 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.The complaint of "o2 tubing found to have a hole in between the nasal prongs" regarding part 1600hf-7-25 was not confirmed.The root cause was not determined.Most likely root cause is molding defect in facepiece.A risk assessment was performed, and the ultimate risk was determined to be low which does not require escalation to the capa review board.There have been no other complaints regarding the same part and a similar issue within the 24 months preceding this reported event.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.Risk assessment: 1.Patient/caregiver performed risk assessment with rma-20017af revision1; the most closely associated risk is id r106 (p1=1, p2=1) with a probability of harm occurring p=1 and severity s=2 which is low risk.The calculated p1 based on complaints and sales in the previous 24 months is 1/5.Therefore, the probability of harm occurring (p) remains the same as the rma and is determined to be low risk per ref-20001-c1 rev2.2.Regulatory/ compliance reviewed ref-20001-c1 rev2, and there is low regulatory/ compliance risk.3.Brand recognition and customer impact reviewed ref-20001-c1 rev2, and there is low brand recognition/customer impact.
 
Event or Problem Description
It was reported that o2 tubing found to have a hole in between the nasal prongs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Common Device Name
NASAL CANNULA (ADULT) HIGH FLOW WITH 7' (2.1 M) SUPPLY TUBE, FOR FLOWS UP TO 15
Manufacturer (Section D)
SALTER LABS
30 spur drive
el paso TX 79906
Manufacturer (Section G)
SALTER LABS DE MEXICO DE S.A. DE C.V.
blvd. independencia #2167 parq
ind. las ameri - hor sur cd ju
chihuahua mexico, mexico 32596
MX   32596
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key23735672
Report Number3004748541-2025-00137
Device Sequence Number9742624
Product Code CAT
UDI-Device Identifier00607411101625
UDI-Public(01)00607411101625
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1600HF-7-25
Device Catalogue Number1600HF-7-25
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Supplement Date Received by Manufacturer11/06/2025
Initial Report FDA Received Date12/05/2025
Supplement Report FDA Received Date02/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-