H11.Additional narrative/data.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
It was reported that a patient with a 27mm 7300tfx mitral valve underwent a valve-in-valve procedure after an implant duration of six (6) years and four (4) months due to severe stenosis.The procedure was performed with a 29mm 9755rsl sapien 3 transcatheter valve.During the valve-in-valve procedure, the first sapien 3 system was inserted, but the operating team was unable to position in the mitral valve.The device was removed (2025-31744-01) and a second s3ur system was inserted in replacement.The procedure was further complicated by patient anatomy and difficulty positioning of second device requiring another removal.The patient expired later that same day due to compartment syndrome, which led to significant blood loss.
|
|
H11: additional manufacturer narrative: updated sections g3, g6, h2, h6 (investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency." there is no evidence to suggest a manufacturing defect caused or contributed to the reported event.There is insufficient information available regarding patient or procedural factors known to cause or contribute to stenosis.Therefore, a definitive root cause cannot be conclusively determined.Per the instructions for use (ifu), structural valve deterioration (svd) is a known potential risk associated with bioprosthetic heart valves.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|