| Model Number |
EM800 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during pre-operative preparation for use, the device was observed to be worn, with rust and corrosion present.The issue was identified prior to patient use.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation determined that the device was tested overheating and the temperature was measured to be 147 degrees fahrenheit.The likely cause of failure was due to corrosion.It was also noted the collet laser markings are faded and the stator is loose.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|