| Model Number |
UHI-3 |
| Medical Device Problem Codes |
Device Alarm System (1012); Leak/Splash (1354); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the high flow insufflation unit was squeaking as if there were an internal air leak.The issue occurred during set up / inspection for use.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h3, h4, h6.The device was returned to olympus for inspection and the reported failure was not confirmed.Based on the results of the investigation, a probable root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information provided by the customer.
|
| |
|