Model Number ANGEL CENTRIFUGE US (REFURBISHED)
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 11/17/2025, a facility representative reported via (b)(4) that an abs-10060r angel centrifuge malfunctioned and would not adequately separate the blood at the end of the cycle during a procedure.The abs-10061t prp kit was unable to be used due to the improper function of the centrifuge, but it did contain blood from the patient who was not affected.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the returned device (if available), pictures, videos, event descriptions, and any additional field data, arthrex concluded the most likely cause of the reported failure.The most likely cause of the reported failure is user error, including incorrect device settings and/or improper surgical technique during use.The complaint allegation was not confirmed.No evidence of that failure was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGEL CENTRIFUGE US (REFURBISHED)
Common Device Name
CENTRIFUGE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23742758
Report Number1220246-2025-05466
Device Sequence Number18729513
Product Code ORG
UDI-Device Identifier00888867249585
UDI-Public(01)00888867249585(10)15086365
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
BK110046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberANGEL CENTRIFUGE US (REFURBISHED)
Device Catalogue NumberABS-10060R
Device Lot Number15086365
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2025
Supplement Date Received by Manufacturer03/06/2026
Initial Report FDA Received Date12/08/2025
Supplement Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-