• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SELF BUNCHING KL 1.8 FIBERTAK, HIP; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SELF BUNCHING KL 1.8 FIBERTAK, HIP; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number SELF BUNCHING KL 1.8 FIBERTAK, HIP
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On november 17, 2025, an arthrex subsidiary employee reported via (b)(4) that two ar-3636h self-bunching kl 1.8 fibertak sutures broke during a labral repair procedure.The first fibertak hip anchor was placed following the standard surgical technique.However, when tension was applied to the suture to perform the knotless technique, the suture unexpectedly broke.To complete the repair, the surgeon requested a second anchor, but the same issue occurred, the suture broke during tensioning despite correct handling and adherence to the recommended procedure.There was no patient harm reported.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF BUNCHING KL 1.8 FIBERTAK, HIP
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23743006
Report Number1220246-2025-05469
Device Sequence Number18757970
Product Code MBI
UDI-Device Identifier00888867362888
UDI-Public(01)00888867362888(11)250324(17)300228(10)15408581
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K251809
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSELF BUNCHING KL 1.8 FIBERTAK, HIP
Device Catalogue NumberAR-3636H
Device Lot Number15408581
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2025
Supplement Date Received by Manufacturer03/05/2026
Initial Report FDA Received Date12/08/2025
Supplement Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-