| Model Number |
TJF-150 |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the duodenovideoscope exhibited the forceps elevator wire had foreign objects.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: it is presumed from the provided information that the foreign material remained for reprocessing was deviated from instructions for use (ifu).The event can be detected/prevented by following the instructions for use which state: ifu states the detection method in tjf type 150 operation manual chapter 3 preparation and inspection.Ifu states the preventative measures in tjf type 150 reprocessing manual chapter 3 cleaning, disinfection and sterilization procedures.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|