| Model Number |
LF1930T |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device experienced a button malfunction.There was no patient involved.Medtronic's initial evaluation of the incident device found the post closest to the activation button was broken off.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted no notable conditions.Functional testing found that the device's jaw opening and closing mechanism was functioning properly.The weld integrity of the device was inspected and was found to be within specification.The device's knife blade advanced and retracted properly.The knife cut of the device was tested on a silicone test strip with acceptable results.The jaw gap of the device was measured and it was found to be within specification.The device was detected when plugged into generator and activated multiple times on a saline soaked gauze pad with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.A rattle was heard from within the device indicating that a piece internally was broken off.Device was disassembled and the post closest to the activation button was broken off.It was reported that the device button malfunctioned.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the sealer divider lf1930t lig jaw thoracic experienced a button malfunction.There was no patient involved.Medtronic's initial evaluation of the incident device found the post closest to the activation button was broken off.
|
| |
|