| Catalog Number |
68-550 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Patient reported reported ¿exchange with no complaint against the device¿ and later healthcare professional reported ¿deflation¿.This record is for the left-side.Device has been explanted and replaced.The device will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a6, b5, d9, h3, h6.Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: deflation: observed an opening through microscopic inspection assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (creases) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Patient reported reported ¿exchange with no complaint against the device¿ and later healthcare professional reported ¿deflation¿.This record is for the left-side.Device has been explanted and replaced.
|
| |
|
Search Alerts/Recalls
|