RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number GB2110X15B
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information that a trilogy evo ventilator was returned to the third-party service center for evaluation of diagnostic test failure on system setup.There was no harm or injury reported.On evaluation, the error logs were downloaded and evaluated.A ventilator service required alarm was confirmed indicating the device software has detected a large pressure drop between the monitor and outlet pressure sensors.The machine flow sensor was replaced to address this issue.Additional findings not related to the malfunction/issue: four-year service was due on the device and was completed.During repair testing, the device failed a test step for fio2 sensor receptacle verification.This test step verifies that the fio2 solenoid is open, and the tubing is not kinked or occluded.This fio2 sensor is used for monitoring purposes only.This would not affect the device's ability to deliver therapy as designed.The fio2 tubing was replaced to address this issue.Re-testing was completed with tests passing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23744094
Report Number2518422-2025-112378
Device Sequence Number8469540
Product Code CBK
UDI-Device Identifier00606959054127
UDI-Public(01)00606959054127
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberGB2110X15B
Device Catalogue NumberGB2110X15B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2025
Initial Report FDA Received Date12/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/29/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-