| Model Number |
LXMJ37L |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total laparoscopic hysterectomy (tlh), the device was difficult or impossible to open.The surgeon was unable to open the jaws, but were subsequently opened after cleaning.It was noted that the device was not applied to the tissue and the knife blade did not protrude beyond the edge of the jaws when the issue occurred.Another device was used successfully with the same generator.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device was difficult or impossible to open.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total laparoscopic hysterectomy(tlh), the device was difficult or impossible to open.The surgeon was unable to open the jaws, but were subsequently opened after cleaning in central sterile supply department (cssd).It was noted thatthe device was not applied to the tissue and the knife blade did not protrude beyond the edge of the jaws when the issue occurred.Another ligasure device was used successfully with the same generator.There was no patient injury.
|
| |
|