Model Number 24MM BASEPLATE, 20° FULL AUG, +2 LAT, ST
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2025, a sales representative reported via email that during the insertion of an ar-9582-30 modular post and an ar-9580-2420-2s baseplate, the two components did not engage properly, resulting in separation.A second set of components, another post and baseplate, was opened to successfully complete the procedure.This was discovered during a reverse total shoulder on (b)(6) 2025.Additional information was received on 11/25/2025: during the procedure, the ar-9582-30 modular post and the ar-9580-2420-2s baseplate disengaged inside the patient after being assembled on the back table.No components broke, and all pieces were retrieved.No other arthrex products were affected.The case was completed using ar-9580-2420-2s modular glenoid system augmented base plate 24 mm+2 (lot 1306512225) and ar-9582-30 modular post for augmented mgs baseplate 30 mm (lot 1749132422).Surgery was delayed by approximately 5¿10 minutes; the additional use of anesthesia is unknown.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided, which may include the returned device (if available), photographs, videos, event description, and any additional field input, arthrex has determined the most likely cause.The most likely cause is user-related, such as incomplete seating or inappropriate torque requirements during device tightening.Dfu-0189-eu-01-eo revision 3 univers revers¿ shoulder prosthesis system.K.Warnings 9.Failure to achieve the appropriate torque requirements when tightening locking screws may result in the premature loosening of the device.The complaint allegation was not confirmed.No evidence of that failure was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
24MM BASEPLATE, 20° FULL AUG, +2 LAT, ST
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23745898
Report Number1220246-2025-05493
Device Sequence Number20309372
Product Code PHX
UDI-Device Identifier00888867352117
UDI-Public(01)00888867352117(11)240901(17)290831(10)1601272339
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K211074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24MM BASEPLATE, 20° FULL AUG, +2 LAT, ST
Device Catalogue NumberAR-9580-2420-2S
Device Lot Number1601272339
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2025
Supplement Date Received by Manufacturer03/27/2026
Initial Report FDA Received Date12/08/2025
Supplement Report FDA Received Date03/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-