This report is being supplemented to provide additional information based on the results of third-party testing, the device evaluation and the complaint investigation.Updated fields: d8, d9, h4, h6, h11.Olympus provided the following result of the culture test, performed at the third-party labs.Sampling date: (b)(6) 2026, sampling from: air/water channel, cfu: 1, bacterial identification: bacillus species.Sampling date: (b)(6) 2026, sampling from: auxiliary channel, cfu: <1, bacterial identification: none.Sampling date: (b)(6) 2026, sampling from: instrument channel, cfu: >80, bacterial identification: bacillus cereus.Sampling date: (b)(6) 2026, sampling from: instrument channel, cfu: 16, bacterial identification: bacillus cereus.Sampling date: (b)(6) 2026, sampling from: suction channel, cfu: >80, bacterial identification: bacillus cereus.Sampling date: (b)(6) 2026, sampling from: air/water channel, auxiliary channel, cfu: <1, bacterial identification: none.Sampling date: (b)(6) 2026, sampling from: instrument channel, cfu: >80, bacterial identification: others.Sampling date: (b)(6) 2026, sampling from: instrument channel, cfu: 13, bacterial identification: bacillus cereus, peribacillus sp.Sampling date: (b)(6) 2026, sampling from: suction channel, cfu: >80, bacterial identification: bacillus cereus.Sampling date: (b)(6) 2026.Sampling from: air/water channel, auxiliary channel, instrument channel, suction channel, cfu: <1, bacterial identification: none.Sampling date: (b)(6) 2026, sampling from: instrument channel, cfu: 2, bacterial identification: staphylococcus warneri.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Despite good faith efforts being made, additional information was not able to be obtained.Based on the results of the investigation, the reported event could not be confirmed, and a root cause could not be determined.Growth of microorganisms were reported through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use before repair, the results conformed to the regulation's recommendation.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|