HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Back to Search Results
Model Number SDA-SYS-3000-2D
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
The user site reported that on (b)(6) 2025, prior to a selenia dimensions mammography system procedure, that the device generated an error code during use.No patient or user injury was reported.A hologic field service engineer (fse) was dispatched to examine the device and check the error code.The fse found that upon reviewing the movement graph of the device, it was observed that the c-arm performed vertical movements without receiving any commands.The fse found that the issue was traced to a malfunction in the motor clutch, which was subsequently replaced.The fse tested the device and verified that the device is now operating according to manufacturer specifications.No further information is currently available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM
Common Device Name
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
600 technology drive
newark DE 19702
Manufacturer (Section G)
HOLOGIC, INC
600 technology drive
newark DE 19702
Manufacturer Contact
ugo chibueze
445 simarano drive
marlborough, MA 01752
MDR Report Key23752609
Report Number3005496266-2025-00108
Device Sequence Number15848876
Product Code MUE
UDI-Device Identifier15420045510616
UDI-Public(01)15420045510616(11)221219
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P010025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSDA-SYS-3000-2D
Device Catalogue NumberSDA-SYS-3000-2D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/08/2025
Supplement Report FDA Received Date06/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/19/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-