Catalog Number 15-265
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported left side "capsular contracture, baker grade iii".The device was explanted and replaced, and will be returned.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture grade 3.
 
Additional Manufacturer Narrative
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.A workmanship was observed through visual inspection not related to the complaint for a split in patch area event.A further investigation is requested.None of the other observations performed during the device analysis creases, deformation and wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Further investigation summary: the review of the documentation associated to the manufacturing process indicates that all devices involved in this work order were released in accordance with abbvie¿s procedures and were no defects/problems/issues found in the documents or in the personnel training related to the reported event.Additionally, no er/ ncr(s) were identified during the investigation associated with this lot and the complaint.According to the device analysis performed in the laboratory was observed split (ss) in patch area, it is out of specification per qa 199.04 the event of capsular contracture unable to observed, and the workmanship has not relation to reported complaint.Therefore, the reported events was not confirmed.A review of the current risk documents pfmea-silicone filled bi and sizer assembly, smooth rev.3.0 was performed, and the event of split in patch is a known event, for which current process controls and inspections are established to reduce the incidence of this defect.Considering that there is no adverse trend for this type of event, no additional actions are deemed required at this time.The issue with split in patch will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.Additional, changed, and/or corrected data: d9, h3, h6.
 
Event or Problem Description
Healthcare professional reported left side "capsular contracture, baker grade iii".The device was explanted and replaced, and will be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key23752845
Report Number9617229-2025-21308
Device Sequence Number21228361
Product Code FTR
UDI-Device Identifier10888628001435
UDI-Public10888628001435
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/29/2014
Device Catalogue Number15-265
Device Lot Number1740374
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2025
Supplement Date Received by Manufacturer02/26/2026
Initial Report FDA Received Date12/08/2025
Supplement Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-