Model Number 1002717
Medical Device Problem Code False Alarm (1013)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
A pump declared an altitude alarm, and the issue did not have an adverse impact on the customer's blood glucose level.The customer¿s insulin was suspended due to the altitude alarm condition, and despite efforts to clean the pump's speaker holes and reconnect the cartridge, insulin could not be resumed.Technical support instructed further troubleshooting steps, and no new cartridge was loaded at that time, with follow-up scheduled to continue addressing the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Common Device Name
ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP
Manufacturer (Section D)
TANDEM DIABETES CARE
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
michael trier
12400 high bluff drive
san diego, CA 92130
858 401-1451
MDR Report Key23753244
Report Number3013756811-2025-275042
Device Sequence Number5865428
Product Code QFG
UDI-Device Identifier00850006613380
UDI-Public00850006613380
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K203234
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1002717
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Initial Report FDA Received Date12/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/18/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TANDEM - CARTRIDGE
Patient Age82 YR
Patient SexFemale
-
-