• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752449
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:(b)(4).
 
Event or Problem Description
A physician reported that the vitrectomy probe did not cut during vitrectomy surgery.The procedure was completed by replacing the pack with new one.There was no patient contact and patient impact.
 
Additional Manufacturer Narrative
Additional information provided in h.6., and h.11.No technical inquiries were received from the customer.A sample was not received at the manufacturing site for evaluation for the report.Therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations, were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.A nonconformance investigation has been completed to investigate the trend identified for probes with would not cut or actuate issues.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23754749
Report Number1644019-2025-05378
Device Sequence Number5928317
Product Code LRO
UDI-Device Identifier00380657524495
UDI-Public(01)00380657524495(17)261231(11)250116(10)1776AU
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752449
Device Lot Number1776AU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/13/2025
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date12/09/2025
Supplement Report FDA Received Date03/10/2026
Date Device Manufactured01/16/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CUSTOM-PAK SURGICAL PROCEDURE PACK.; CUSTOM-PAK SURGICAL PROCEDURE PACK.
Patient SexUnknown
-
-