|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
|
It was reported that, during a real intelligence cori tka surgery, the coriograph pre-op services plan indicated a size 5 j2 bcs femur, with planned posterior resections of 9.0 mm medially and 7.5 mm laterally, set at 0° referencing the pca.However, after registration, the initial posterior cuts displayed were 12.5 mm medially and 7.5 mm laterally, also at 0° referencing the pca.Due to this discrepancy, it was verify using the manual sizing guide.The guide confirmed a size 5 femur.The procedure was resumed, as normal, after a non-significant delay.No further complications were reported.
|