| Model Number |
DDMB1D4 |
| Medical Device Problem Codes |
Failure to Charge (1085); Electrical /Electronic Property Problem (1198); Delayed Charge Time (2586)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) experienced a charge circuit timeout due to a charge time over twenty three seconds.The icd had reached end of service (eos) and was explanted and replaced. in addition, there were short v-v intervals on the right ventricular (rv) lead. the rv lead remains in use. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected: 6b medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient went into torsades and received forty shocks.The physician believed that the torsades was due to r-on-t.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|