| Model Number |
K173 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.This device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.This device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis; therefore, a technical analysis could not be performed.A risk review was completed and confirmed that the event of message - error code 1003 was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of unable to exclude device problem.
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.This device remains in service.No adverse patient effects were reported.
|
| |
|