| Model Number |
990063-020 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pericardial Effusion (3271)
|
| Date of Event |
11/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the 990063-020 achieve mapping catheter with lot 230631164 was returned and analyzed.Visual inspection of the loop segment area showed the loop was intact with no apparent issues.No damage was observed along with the tip/loop section of the mapping catheter.Visual inspection of the pebax tubing area showed the pebax tubing was intact with no apparent issues.No damage was observed along with the pebax tubing section of the mapping catheter.Visual inspection of the electrode(s) showed the electrode(s) was intact with no apparent issues.All electrodes exist on the loop section and no cosmetic issue or anomalies were identified.Visual inspection of the shaft segment area showed that the shaft was broken approximately 10.5 inches from the lemo connector.No electrocardiogram (ecg) signal wires were broken.Visual inspection of the introducer showed the introducer was intact with no apparent issues.No damage or any other issue was observed along with the introducer.Visual inspection of the lemo connector showed the lemo connector was intact with no apparent issues.No damage or any other issue was observed along with the lemo connector.The functional test was performed using a multimeter.The mapping catheter was connected to the test cable.The continuity and impedance measurement between the electrodes and the other side of the cable showed the electrode's continuity and impedance to the cable are normal.In conclusion, the reported pericardial effusion cannot be assessed through product testing.The mapping catheter failed the returned product inspection due to a broken shaft.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id: 990063-020, product type: mapping catheter.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cardiac ablation procedure, a patient developed a pericardial effusion.A pericardiocentesis was performed.The patient was stabilized in the electrophysiology lab before being transferred to the intensive care unit for recovery with the drain in place.The patient required unexpected hospitalization.The procedure was completed with cryo.No further patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|