RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number CA2110X12B
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging the touchscreen was damaged on a trilogy evo device.The device was not in use on a patient at the time of the occurrence.The trilogy evo device was returned to a third-party service center for evaluation, and the customer¿s complaint was confirmed.During the evaluation, the third-party service technician did not provide any findings on the cause of the damage display for this device.In conclusion, the device's display screen was replaced to address the issue.Additionally, during the service of the device, the detachable battery pack, muffler assembly, and air inlet foam filter were replaced for preventative maintenance unrelated to the reportable issue.The particle filter, system printed circuit assembly tubing, blower chassis tubing, fio2 tubing, and low flow o2 tubing were replaced for contamination unrelated to the reportable issue.The device passed all final testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23763744
Report Number2518422-2025-112472
Device Sequence Number10091240
Product Code CBK
UDI-Device Identifier00606959052307
UDI-Public(01)00606959052307
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCA2110X12B
Device Catalogue NumberCA2110X12B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2025
Initial Report FDA Received Date12/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-