| Catalog Number |
0167LT20 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter was placed after the operation, but the balloon ruptured two days later and was removed naturally.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed ¿ cause unknown.The reported failure was able to be reproduced.The product was used for urological care.Visual inspection noted sac burst along lumen without missing pieces.Introduced water into the inflation lumen and did not experience any obstructions to impede flow.Microscopic examination found no conditions that could be associated with the reported event.The exact cause on how and when problem occurred could not be determined.The potential root cause for this failure could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/operator error).The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.Therefore, no additional action required at this time.Correction: d, f,h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter was placed after the operation, but the balloon ruptured two days later and was removed naturally.
|
| |
|
Search Alerts/Recalls
|