|
MEDTRONIC SOFAMOR DANEK USA, INC. CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
| Model Number |
9010000849 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter details are unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, user facility) regarding a device used in an unknown sp inal therapy.It was reported that the inserter was broken.There was no patient involved.
|
| |
|
Search Alerts/Recalls
|
|
|