| Catalog Number |
MF-410225 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "implant rupture" and "exchange from textured to textured to smooth due to the patient concern of the implant".This record is for the left side.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b3, b6, g2.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "implant rupture" and "exchange from textured to textured to smooth due to the patient concern of the implant".This record is for the left side.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "implant rupture" and "exchange from textured to textured to smooth due to the patient concern of the implant".This record is for the left side.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: rupture: observed an opening through microscopic inspection assessed as unidentified (tear) opening.No further actions are required as it is not related to the manufacturing process.Anxiety - product/ procedure: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (creases) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Search Alerts/Recalls
|