• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPACELABS HEALTHCARE SPACELABS HEALTHCARE QUBE COMPACT MONITOR; QUBE BEDSIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPACELABS HEALTHCARE SPACELABS HEALTHCARE QUBE COMPACT MONITOR; QUBE BEDSIDE Back to Search Results
Model Number 91389
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that a qube mini bedside monitor had shut off on its own while monitoring a patient.There was no patient or user harm associated with this event.The device was returned to spacelabs healthcare for further evaluation.The equipment service center (esc) technician evaluated the device and was able to verify the reported issue.The esc technician found that the unit had a faulty cpu pcba.The faulty board was replaced and after passing all functional and performance testing, the device was returned to the customer.The cause of the reported issue was due to a faulty cpu pcba.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS HEALTHCARE QUBE COMPACT MONITOR
Common Device Name
QUBE BEDSIDE
Manufacturer (Section D)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer (Section G)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer Contact
whitney snyder
35301 se center st
snoqualmie, WA 98065
MDR Report Key23769637
Report Number3010157426-2025-00229
Device Sequence Number9687446
Product Code MHX
UDI-Device Identifier10841522100239
UDI-Public(01)10841522100239(11)180131
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K120616
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 12/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number91389
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2025
Initial Report FDA Received Date12/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-