• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ NMIC-306; SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ NMIC-306; SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION Back to Search Results
Catalog Number 449292
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using the bd phoenix¿ nmic-306 a patient isolate (escherichia coli) had a high mic for the drug ceftolozane/tazobactam (resistant) while the drug meropenem was susceptible and minocycline was intermediate.In addition, the extended-spectrum beta-lactamase (esbl) marker was flagged.The user noted performing repeat testing.No health impact or consequence reported.
 
Additional Manufacturer Narrative
There were multiple 510k numbers reported to be involved.The information for the additional 510k is as follows: g4.Pma / 510(k)#: k020322, k022129, k023444, k023634, k023858, k024153, 031530, k031699, k031912, k032299, k032567, k032655, k032675, k033362, k033458, k033558, k033560, k041384, k042932, k052269, k060214, 060217, k060257, k060444, k060447, k061327, k061355, k061867, k062207, k062944, k063301, and k06348.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields have been updated with corrected and/or additional information: d10.Device available for eval- yes.D10.Returned to manufacturer on 12-dec-2025.H3.Device eval by manufacturer- yes.Investigation summary: this complaint is for high mic ceftolozane-tazobactam (ct) when using phoenix panel nmic-306 (catalog number 449292) batch number 5266643.The customer returned panels, isolates, phoenix generated lab reports and binary files for investigation.The lab reports show high mic results for ct with escherichia coli and citrobacter braakii when using the complaint batch.The three isolates received were c.Braakii 4054, e.Coli 1357 and e.Coli 2641.The batch history record review for the complaint batches was satisfactory, and no quality notifications were generated during manufacturing or inspection.Formulation, filling, and packaging processes were within specifications.In process checks are performed during manufacturing at designated intervals per procedures.Qc inspection and testing were satisfactory at time of release.Complaint trending was performed, and no trends were identified associated with this defect.To investigate, customer returned panels of the complaint batch were inoculated with customer returned isolates e.Coli 1357, e.Coli 2641 and c.Braakii 4054 and placed in a phoenix m50 to evaluate for ct mic results.Next, control panels of the same material but different batch were also inoculated with customer returned isolates e.Coli 1357, e.Coli 2641 and c.Braakii 4054 and placed in a phoenix m50 to evaluate for ct mic results.The customer returned and control panels tested with customer returned isolates e.Coli 1357 and e.Coli 2641 returned susceptible ct along with susceptible mem results.The customer returned and control panels tested with customer returned isolate c.Braakii 4054 returned susceptible mic results for mem along with resistant mic results for ct.Bd r&d reviewed the customer's results side by side with the results from bd's internal complaint testing.It was determined through review of customer results that isolate # 1 (e.Coli accession # 8770) and isolate # 2 (c.Braakii 4054) likely possess an ampc resistance mechanism due to resistant results for cefazolin, ceftazidime, ceftriaxone and cefoxitin along with susceptible and intermediate results for cefepime.It is to be noted that the presence of ampc can cause ct resistance but would not impact carbapenems.Therefore, isolates #1 and #2 were determined to be a strain specific failure and not indicative of the phoenix system as a whole.Analysis of the phoenix growth plots for isolates # 3 (e.Coli 1357) and isolate # 4 (e.Coli 2641) indicates low level contamination which could be attributed to introducing multiple organism strains into the id broth inoculum during the colony picking process, bioburden within the ap system or environmental issues.For further review, complaints due to high mic for ct across this material (#449292) were reviewed and there are no trends associated with this defect.As no trends were identified, no corrective actions are slated at this time per bd procedures.Bd will continue to monitor for trends and take action as necessary.Please continue to communicate any additional concerns.
 
Event or Problem Description
It was reported while using the bd phoenix¿ nmic-306 a patient isolate (escherichia coli) had a high mic for the drug ceftolozane/tazobactam (resistant) while the drug meropenem was susceptible and minocycline was intermediate.In addition, the extended-spectrum beta-lactamase (esbl) marker was flagged.The user noted performing repeat testing.No health impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ NMIC-306
Common Device Name
SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key23770024
Report Number1119779-2025-05487
Device Sequence Number9742799
Product Code LON
UDI-Device Identifier30382904492923
UDI-Public(01)30382904492923(17)260919(10)5266643
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
SEE H11
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/19/2026
Device Catalogue Number449292
Device Lot Number5266643
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2025
Supplement Date Received by Manufacturer01/22/2026
Initial Report FDA Received Date12/10/2025
Supplement Report FDA Received Date02/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-