| Catalog Number |
305617 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd syringe 20ml ll tip conv pak leaked it was reported by the customer that another leaky syringe issue.Verbatim: rcc received a complaint via email.Email(s) attached i am reaching out with another leaky syringe issue.We have four more syringes with fluid that got behind the plunger.The event happened (b)(6) 2025 while the technician was drawing up antibiotic.Would you like these syringes as well?.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: our quality engineer inspected the 4 physical samples submitted for evaluation.The reported issue of leakage other was not confirmed upon inspection of the samples.Analysis of the samples resulting in all samples passed a leak test.Bd could not determine a manufacturing related root cause since the reported defect was not confirmed during the evaluation of the sample.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this batch meets our manufacturing specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Search Alerts/Recalls
|