This follow-up report is being submitted to relay additional information.The following sections were updated: b4,b5,d2,e1,e2,e3,g1,g3,g6,h1,h2,h3,h4,h6 the cmp was confirmed -visual evaluation of the provided photos/returned product found the sterile packaging exhibits damage that is visually consistent with thermal damage.Sterility is considered not breached.-review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.-reported event is not related to a combination of products; therefore, a compatibility review is not applicable.-review of complaint history identified additional similar complaints for the reported part and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.-medical records were not provided.-the root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|