It was reported that patient had blakemore placed at 0152 inside of patient with 250ml of air inserted into fatric port by a physician.At 0344, patient was found bleeding from mouth, and it was found by physicians that the gastric part of the blakemore was ripped while in the patient.A new blakemore was placed, and patient was transported defective blakemore was placed and patient was transported defective blakemore was given to sterile processing.No medical intervention was reported.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation the reported event is addressed within the labeling as it states: warning: it is mandatory, assuming that the balloon has controlled the bleeding, to determine the state of the liver prior to removal of the esophageal balloon.Removal of the balloon too early in an individual with marked prothrombinemia or thrombocytopenia will lead to resumption of bleeding.It is safe to tampon esophageal varices for prolonged periods, months, if necessary, in preparation for operation if no traction is applied.1.Test balloons for air leaks and coat the lower parts of the tube and both balloons with a thin coat of lubricating jelly.2.Evacuate all the air from the gastric and esophageal balloons and reinsert the rubber plugs.3.After spraying the nostrils with a topical anesthetic, pass the tube through the nostril until the tip is in the posterior pharynx or throat.As the patient swallows water, pass the tube to at least the 50 cm mark so that the tube numbers marked on the tubing will be along the right lateral aspect of the esophagus.Proper placement may be verified by x ray.Inflate the gastric balloon with 200cc to 250cc of air, clamp conical portion of the gastric balloon inflation tube approximately 3cm from the end (avoid the small tubular portion of the gastric balloon), and pull back on the tube until resistance is encountered.(the gastric balloon is so constructed that with the numbers on the tube in the right lateral position, the cardia and fundus of the stomach will be engaged.) after the gastric balloon has been engaged, with a minimum of tension on the tube, fix the upper end of the tube as it emerges from the nostril by a cuff of sponge rubber held in place by an adhesive tape band.Finally, inflate the esophageal balloon to 35 to 40mm of mercury pressure.4.Next, aspirate through the stomach tube all air, water and blood.It is necessary during this step to irrigate the tube and stomach continuously with 50cc of water to prevent clotted blood from plugging the tube.Caution: it is most important that the stomach be cleared of blood promptly since subsequent sampling will be a true index of the effectiveness of hemorrhage control from the varices in the esophageal wall.5.If, after careful lavaging of the stomach for 30 minutes bright red bleeding continues, increase the esophageal balloon pressure to 45 mm of pressure.At the same time, continue to wash the stomach with water so that the moment bleeding is controlled, it will be detected.Under normal circumstances esophageal balloon pressure should not exceed 45 mm of mercury, and it is not uncommon for a patient to complain of substernal pressure as the higher pressures in the balloon are reached, particularly if they are reached too rapidly.When the balloon is in the proper position, the pressure will vary with respiratory movements and esophageal contractions.Occasionally these variations will cause the balloon pressure to rise to 70mm of mercury.The base line pressure is the one of importance, and not the transient peak pressures.6.If esophageal balloon pressure is maintained at 45 mm of mercury and bleeding continues, it is suggestive of a gastric wall varix.In this case, the patient is given additional sedation if necessary and the tube is snugged up on the nose firmly and taped securely.Finally, the gastric balloon is inflated with more air gradually up to a total of 300 to 400cc of air.Warning: this degree of pressure causes ulceration of the mucosa within a few hours; therefore, patient must be prepared for suture of the gastric varices as soon as the general condition permits.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Corrections: d, e udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that patient had blakemore placed at 0152 inside of patient with 250ml of air inserted into fatric port by a physician.At 0344, patient was found bleeding from mouth, and it was found by physicians that the gastric part of the blakemore was ripped while in the patient.A new blakemore was placed, and patient was transported defective blakemore was placed and patient was transported defective blakemore was given to sterile processing.No medical intervention was reported.Per customer response received on 22dec2025, no patient harm was noted.
|