| Catalog Number |
D134801 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cardiac ablation procedure with a thermocool smarttouch sf and after performing pulmonary vein isolation and cavotricuspid isthmas ablation (3 hours after the patient entered the room) the medical team became unable to flush the catheter.This specifically occurred at the time of ablating the gap area of the cavotricuspid isthmas.There were no error codes preceding the issue.Char was suspected but no char was visually confirmed.The issue was resolved by replacing the catheter with another one.After that, the procedure was completed without any problems.
|
| |
|
Additional Manufacturer Narrative
|
|
On 26-dec-2025, the bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.It was reported that a patient underwent a cardiac ablation procedure with a thermocool smarttouch sf and after performing pulmonary vein isolation and cavotricuspid isthmas ablation (3 hours after the patient entered the room) the medical team became unable to flush the catheter.This specifically occurred at the time of ablating the gap area of the cavotricuspid isthmas.There were no error codes preceding the issue.Char was suspected but no char was visually confirmed.The issue was resolved by replacing the catheter with another one.After that, the procedure was completed without any problems.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.An irrigation test was performed, and the device was irrigating correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device 31652121l number, and no internal actions related to the reported complaint condition were identified.The irrigation issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the use of the device that may have affected its performance.The instructions for use (ifu) contain the following information: before use, verify that the irrigation ports are patent by infusing heparinized normal saline through the catheter and the irrigation tubing.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|