MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3501635 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
09/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.However, a photo of the suspect medical device was provided.An evaluation using the image was completed.Upon visual evaluation of the image provided in the complaint, the implant was observed deflated.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 250cc mentor smooth round moderate profile saline breast implant prosthesis.Post-operatively, the patient was diagnosed with a deflated left implant during a physical exam.As a result, the patient underwent left-sided exchange with a mentor saline implant on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
On january 14, 2026, the mentor analysis lab received the suspect medical device for evaluation.Paperwork received with the device provided a patient identifier.Patient identifier: (b)(6).On february 10, 2026, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted a visual inspection, microscopic examination, and shell thickness of the device.Visual analysis of the returned device revealed that the breast implant was found to have a separated material on the posterior view, measuring 1.4 cm x 1.1 cm.In addition, the device was received in two (2) parts.Microscopic examination was performed, and the cause of the tear could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.Manufacturer¿s reference number: (b)(4).
|
| |
|