| Model Number |
HX55V-S0W |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
11/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
User reported a blood to glycol leak during the case.No adverse consequences occurred for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information related to the event will be available after completion of investigation activities, which would be addressed in a follow-up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Preliminary findings available in current follow-up 001 report: visual inspection and defect identification: two symmetrical holes were found in the central area of the device where the knurled pattern changes direction.Scanning electron microscopy (sem): sem imaging was used to characterize the dimensions of the holes and to obtain details of the metal sheet surface near the holes.Fluid test results: testing confirmed that even with the holes present, under worst-case conditions, flow occurs from blood to htf, ensuring that no htf can enter the blood compartment.This is due to the frosty design: htf is drawn out of the heat exchanger, so the pressure inside the htf compartment is always negative, while the blood compartment is always under positive pressure.As a result, the pressure differential across the metal sheet guarantees that only blood can enter the htf compartment, not vice versa.Further testing is ongoing to gather more details that will help identify the root cause of the problem.These tests include: analysis of metal sheets performed by an external lab; study of current/voltage versus hole formation.A final and complete report will be released when all the results will be available.
|
| |
|
Event or Problem Description
|
|
User reported a blood to glycol leak during the case.No adverse consequences occurred for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Preliminary analysis performed revealed the presence of one hole in the metal sheet.Nevertheless, the root cause of this issue has not been identified.Additional investigations are ongoing.Therefore, an additional follow-up communication will be done after completion of investigation activities.
|
| |
|
Event or Problem Description
|
|
User reported a blood to glycol leak during the case.No adverse consequences occurred for the patient.
|
| |
|
Event or Problem Description
|
|
User reported a blood to glycol leak during the case.No adverse consequences occurred for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the investigation performed, it was confirmed that blood to htf leakage is generated by holes present in the hx55 metal sheet.By combing the results of the investigation, it is possible to state that given the inherent complexity and timescale of the spontaneous corrosion mechanism, hole formation within the timeframe reported by the customer (minutes to hours) would only be plausible in the presence of external accelerating factors that cannot be predictable a priori.Nevertheless, considering the following: for what concern the potential impact on patient safety, even with the holes present in the metal sheet, under worst-case conditions, flow occurs from blood to htf, ensuring that no htf can enter the blood compartment.This is due to the quantum heater-cooler machine design: htf is drawn out of the heat exchanger, so the pressure inside the htf compartment is always negative, while the blood compartment is always under positive pressure.As a result, the pressure differential across the metal sheet guarantees that only blood can enter the htf compartment, not viceversa.This represents a safety mechanism acting to protect the patient in case of metal sheet loss of integrity.The instructions for use clearly indicate that the priming procedure must be completed shortly before device use, since prolonged contact with the priming solution may compromise the device's integrity and performance.The malfunctioning of the metal sheet is already considered as potential hazard in the heat exchanger product risk analysis with a benefit-risk ratio still acceptable after implementation of all risk control measures.The actual incidence rate of this kind of issue is currently < 0,1% (0,019%) compared to the estimated value in the risk analysis (<1%) considering the units sold worldwide.For these reasons, no further actions are deemed required.
|
| |
|