• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number DDU-120A
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that their aed's asi is flashing red, and the aed will not power on.The customer did not report this occurring during patient use.
 
Additional Manufacturer Narrative
Analysis of the log files from the aed is ongoing and the cause of the complaint is not known.
 
Additional Manufacturer Narrative
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.The cause of the aed not powering on was related to the customers failure to maintain the battery pack.As a follow up with the customer, the proper maintenance of this device was reviewed.Once the new battery was installed and a manual self test performed the aed functioned as designed and was able stay in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford CT 06405
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key23773878
Report Number3003521780-2025-02020
Device Sequence Number21173509
Product Code MKJ
UDI-Device Identifier10815098020048
UDI-Public(01)10815098020048
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-120A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/10/2025
Supplement Report FDA Received Date01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-